Placed a hand held rad monitor near the faraday cup and took rate readings every 20 seconds, right after inserting the gate valve in front of SEC:
seconds counts/s
0 590
20 499
40 474
60 400
80 353
100 290
120 259
140 234
160 207
180 187
200 175
220 146
240 135
260 124
280 120
300 109
320 107
340 103
360 85
380 86
400 84
420 94
440 81
460 76
480 76
500 74
Then fitted this for half life.
Assuming a steady state background of 70 cps, and using the first 7 data points... the half life comes out as 82.1 +/-3.9 s.
Excellent agreement with known halflife of 59Cu.
|