<p><strong>Thursday, 7th September:</strong></p>
<p>Vaccum after pumping all night preamps off, chiller off</p>
<ul>
<li>09:37: IMG1 2.3E-5 mbar, Chamber 7.5 E-5 mbar; Julabo off;</li>
<li>09:38: IMG1 2.1E-5 mbar; Chamber 6.6E-5 mbar; Julabo 20 Celsius</li>
<li>09:41: IMG1 2E-5; Chamber 6.6E-5; Julabo 5 Celsius</li>
<li>09:55: IMG1 1.3E-5; Chamber 1.3E-5</li>
<li>10:10 IMG1 1.2E-5; Chamber 4.5E-5;</li>
</ul>
<p>Preamp power ON, Chiller on 5 Celsius</p>
<ul>
<li>10:25 IMG1 1.3E-5; 5E-5; Thermocoupling Readout 16 Celsius</li>
<li>11:20 IMG1 1.5E-5; 6.1E-5 Thermocoupling Readount 19.3</li>
<li>13:08 IMG1 1.6E-5; 5.5E-5</li>
<li>14:16 IMG1 1.4E-5; 5.1E-5</li>
<li>15:40 IMG1 1.4E-5; 4.9E-5</li>
<li>17:55 IMG1 1.3E-5; 4.5E-5</li>
</ul>
<p>Preamp OFF, chiller off</p>
<p><strong>8th September</strong></p>
<ul>
<li>1.2E-5; 3.9E-5</li>
</ul>
<p>--> Leak test with He, replace target ladder with blank and refit feedtrhrough (upstream detectors), identify some scratches</p>
<p>start pumping</p>
<ul>
<li>23:50 IMG1 2.2E-5; Chamber 9E-5</li>
</ul>
<p><strong>9th September</strong></p>
<ul>
<li>08:17 IMG1 1.3E-5; 5.1E-5 mbar</li>
</ul>
<p>Leak test but no leaks found anywhere now.</p>
<p>Chiller on 5 Celsius, preamplifiers ON</p>
<ul>
<li>10:55 IMG1 9.1E-6; 3.8E-5</li>
<li>12:00 IMG1 8.9E-6; 4.0E-5</li>
<li>14:50 IMG1 8.1E-6; 3.6E-5</li>
</ul>
<p>Preamplifiers OFF</p>
<ul>
<li>15:05 IMG1 6.3E-6; 2.9E-5 (9.8 C)</li>
<li>15:12 IMG1 6.1E-1; 2.9E-5 (9.3 C)</li>
<li>15:25 IMG1 5.9E-6; 2.8E-5 (9.0 C)</li>
</ul>
<p>Chiller OFF (not sure when exactly chiller was turned off)</p>
<p><strong>Sunday 10/09/2017</strong></p>
<ul>
<li>22:50 IMG1 8.8E-06; 3.4E-05</li>
</ul>
<p><strong>Monday 11/09/2017</strong></p>
<ul>
<li>9:20 IMG1 8.2E-06; 2.7E-05</li>
</ul>
<p>Leak test on Cooling loop. --> Big leak detected. Open up chamber to replace cooling.</p>
<p><strong>Tuesday 12/09/2017</strong></p>
<p>Several tests done for cooling, it seems that even the smallest possible cooling loop is leaky.</p>
<ul>
<li>19:00, close chamber and start pumping</li>
<li>19:13 start both turbo pumps at 9.9E-1 mbar</li>
<li>19:23 P=2.7E-4 mbar</li>
</ul>
<p>Note that big Turbopump has been switched to normal start so pumping times not comparable with before. Maximum wattage of pump should be 420 W</p>
<p><strong>Wednesday 13/09/2017</strong></p>
<ul>
<li>9:15 8.5E-6 mbar; 3.1E-5mbar</li>
<li>9:30 Cooling ON, switched to -10C</li>
<li>10:05 4.4E-6 mbar; 1.7E-5 bar (-9.9C on chiller)</li>
<li>10:15 preamps on, -10 C, 4.1E-6; 1.6E-5 mbar</li>
<li>10:22 T readout 5.6 C, 4.6E-6; 1.8E-5</li>
<li>10:26 T readount 6.2 C; 4.8E-6 mbar, 1.9E-5 mbar</li>
<li>10:40 4.9E-6 mbar; 1.9E-5 mbar</li>
</ul>
<p>Preamps off, cooler off</p>
<p>open chamber for detector installation and some checks; close chamber and pump</p>
<ul>
<li>18:55 7.2E-5; 3.3E-4 mbar</li>
<li>20:45 4.4E-5; 1.8E-4 mbar (nopreamp, no chiller)</li>
</ul>
<p><strong>Thursday, 14 September 2017</strong></p>
<ul>
<li>9:25 1.1E-5mbar; 4.3E-5 mbar</li>
</ul>
<p>open chamber to discuss with beam operators</p>
<ul>
<li>11:05 close chamber start pumping</li>
<li>11:20 1.2 mbar, turbo pumps on</li>
<li>11:32 4.1E-4 mbar; 1.1E-3 mbar</li>
</ul>
<p> </p> |