Entry time:
Wed Mar 26 10:23:36 2025
Author
*
:
Subject
*
:
Notes from an email: Tested the chamber with the leak checker. We got 3.7 x 10^-6 mbar liters / sec. I estimate the chamber's leak rate around 1 mbar / 4 hours or so, and I randomly guess its volume is 100 ml. If I smash those numbers I get ~10^-5 mbar liters / sec which is a similar order. I turned off the pumps for the weekend but left the He gas in the cell.