| MONNET |
 |
|
|
Message ID: 8
Entry time: Fri Feb 27 12:41:34 2026
|
| Author: |
SB, TD, FGI |
| Subject: |
Friday 27 February 2026 |
|
|
BNC PB-5 settings
Amplitude 5.0V
Attenuator x10
Decay time 1ms
Polarity +
Frequency 20Hz
Delay MIN (250ns)
Silena 7710 Quad Bias
Channel 1 connected to preamp A, channel 2 to preamp B
DSSSD#1 (preamp A) bias -70V I_L -0.11uA
DSSSD#2 (preamp B) bias -70V I_L -0.10uA
CAEN V560 Scaler
Channel
0 Triggers
1 Triggers accepted (STFC DL SAC Mon 2 output )
2 Clock (1kHz)
3 Pulser (positive NIM logic output -> negative NIM output via Phillips 794)
4
:
:
9 OR DSSSD#1 p+n strips 0-15
10 OR DSSSD#1 n+n strips 0-15
11 OR DSSSD#2 p+n strips 0-15
12 OR DSSSD#2 n+n strips 0-15
CAEN V560 scaler channels 0-4 read per event
DAQ Trigger
DSSSD#1 n+n strips OR DSSSD#2 p+n strips OR DSSSD#2 n+n strips (LeCroy 429A)
DAQ trigger re-shaped from c. 100ns to c. 1us by Phillips 794
Logic cable #4 appears to kill 48-way OR logic module - workaround - cable 4 disconnected - scaler #9 no reading
Silena 9418/6V ADC settings
RTP 0 (2us)
LLD 6 (48 channels)
No ADC data from 3 channels
S9418#1.adc14
S9418#2.adc8
S9418#2.adc9
All other channels appear to be OK
Pulser peak width c. 7 ch FWHM p+n strips, c. 8 ch FWHM n+n strips
See attachments 8-11
RAl109 amplifier output DSSSD#1 p+n strip 0 - attachments 1, 2 and 4 - beam OFF, no significant extrinsic noise
Photograph of experiment electronics and sample chamber - attachments 3, 5-7 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|