AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
nTOFCapture |
 |
|
Message ID: 31
Entry time: Thu Jul 13 12:27:44 2023
|
Author: |
Michael Donnachie, CLW |
Subject: |
Calibration Parameters |
|
|
Calibration parameters for each detector:
detector |
y-intercept (value) |
y-intercept (error) |
gradient (value) |
gradient (error) |
1 |
|
|
|
1.56343E-06 |
2 |
|
|
|
1.96751E-06 |
3 |
|
|
|
1.33883E-06 |
4 |
|
|
|
1.22941E-06 |
A linear function was fitted to the reference point channel (from experimental histogram) and reference point energy (from simulation histogram). The reference point 'r' was r=mean+(FWHM/2) for a gaussian fitted to each compton edge.
The comparrisons between the calibrated experimental and simulation histograms has been attached. P1 (left) and P2 (right) on the Y88 hisograms indicate whether the histograms were scaled with respect to the leftmost or rightmost compton edge (only to allow easier comparrison).
Broadening parameters applied to the simulations
float a1 = 0.004947965;
float b1 = 0.005227974;
float a2 = 0.007163802;
float b2 = 0.007549987;
float a3 = 0.001256462;
float b3 = 0.006410511;
float a4 = 0.003907402;
float b4 = 0.004033837;
sigma= sqrt(axE+bxE^2) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|