<p>There seems to be issues with the EAR2 C6D6 T histograms. For Au the earlier run seems to be shifted for detector 2 compared to the rest. </p>
<p>For Si plotted is 3 different runs 216123, 216128 and 216135 for both detectors. The histogram is messy but highlights that there is a problem. The single Si resonace should be around 20 000ns and this is only the case for run216128. </p>
<p>Apologies for no legend on plots. </p>
<p>Au22<br />
det1 run216158 black<br />
det1 run216109 red<br />
det2 run216158 green<br />
det2 run216109 blue</p>
<p>Au22<br />
det1 run216157 black<br />
det1 run216108 red<br />
det2 run216157 green<br />
det2 run216108 blue<br />
</p> |