Message ID: 41
Entry time: Fri Jul 28 12:19:10 2023
|
Author: |
Michael Donnachie |
Subject: |
EAR2 consistency checks, PKUP BCT SILI and normalised counts |
|
|
The EAR2 ratios of BCT/PKUP, BCT/SILI and PKUP/SILI plotted for each run for samples Si, Sinat, Au20, Au22 and Dummy.
Also included is the normalised counts (counts/BCT) with 7 detectors on one plot (STED8 was ommited). 'counts' for Si was integreated over the Si resonance, for Au it was integrated over the largest resonance. For Dummy and Sinat the counts were integrated over a large range of 1e+4 to 1e+6ns. The Sinat resonance was too small for adequate statistics to only integrate over the resonance hence why a large range was used. The C6D6 normalised counts were not plotted as there seems to be issues with this detector in EAR2.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|