Entry  Fri Sep 9 10:01:18 2016, Sarah, Leakage currents 09.09.16 
EFED 0.04uA
DEED 0.05uA
PHDI 0.13uA
Entry  Tue Aug 9 15:57:52 2016, Sarah, Leakage currents 09.08.16 
 Morning - 9am
E= 0.03uA
dE= 0.04uA
Entry  Mon Aug 8 11:02:13 2016, Sarah, Leakage currents 08.08.16 
 E 0.03 uA
dE 0.04 uA
PHDI 0.05 uA
Entry  Fri Sep 2 12:01:57 2016, Sarah, Leakage currents 02.09.16 
E 0.03
dE 0.05
PHDI 0.12
Entry  Wed Aug 3 14:48:09 2016, Sarah, Leakage currents 02.08 - 03.08 
 02.08:
dE = 0.03 uA
E = 0.03 uA
Entry  Thu Oct 6 09:28:33 2022, Claudia, Items to be replaced for next run 20221005_122654.pdf
Missing/broken items identified

- long rails for fixing chamber on support

- vacuum valve (ours is broken)
Entry  Thu May 26 09:20:32 2022, Nick / Claudia, Inventory of Items at n_TOF for (n,cp) experiments rn_image_picker_lib_temp_e85b2099-7b73-4246-843d-6351800c2709.jpgrn_image_picker_lib_temp_3ff5f819-1e0c-4278-bbbc-01038a0a0c94.jpg
Detectors currently in the boxes:

2x MSL type W1(SS)-50, 2902-4 (53um), 2940-6 (55um)

1x MSL type W1(SS)-20 2837-33 (20um)
Entry  Thu Aug 10 10:05:48 2023, CLW, Gafchromics Foil 20230809_165433.jpg20230810_085457.jpg20230809_165436.jpg
Gafchromics foil was mounted on empty brass frame back to back with Li sample

Neutron beam goes well through the sample. Centered ~1 cm from sample center. Conclusion: no changes needed (too risky to make it worse and it
is ok now)
Entry  Wed Jul 27 13:20:53 2016, Sarah, Gafchromic images 6x
 
Entry  Mon Jun 23 09:36:32 2025, CLW, Expected Spectra for Li and B in a 50 um Si detector 50_um.png
B looks as we are used to , Li tritons are now stopped in the detector and should appear around 2.8 MeV. 
Entry  Mon Oct 2 08:53:06 2023, Emmanuel, Empty Frame Empty_Frame.jpeg
An empty frame (on the top) was placed on Friday night.
Entry  Wed Aug 3 09:48:31 2016, Sarah, EFED10 alpha calibration checks 
 

    
        

           
Entry  Fri Aug 11 16:50:58 2023, Nikolay Sosnin, EDET24 EDET24_Card.JPGEDET24_Cable.JPG
After inserting aluminium target, previously-dead channel 24 on EDET was re-plugged into a 12-bit card SPD02872. It is the only EDET on 12-bit card,
the rest are on 14-bit. From pulsers, there appears to be a 90 ns offset from this EDET to others, with EDET24 preceeding the rest. The last channel on
14-bit cards was broken, hence havng to use 12-bit.
Entry  Thu Sep 12 15:26:30 2024, Nick, EAR2 Periodic Noise IMG_20240912_161937023_HDR.jpg
 
Entry  Fri Aug 11 13:16:56 2023, Nikolay Sosnin, E Front Incorrect Cabling 
E front 2x16-way to 1x34-way IDC cable from the Junction Box to the IDC-Lemo 00 adaptor was found to be inserted upside-down - the 16-way IDC connectors
did not have polarisation keys - meaning the E detector p+n junction strip signals collected with 26Al for the first night are not working. Signals
from the dE p+n junction strips and E n+n Ohmic strips will be OK. The mistake occurred at the conclusion of the noise tests yesterday - see https://elog.ph.ed.ac.uk/nToF/52
Entry  Tue Sep 5 16:09:45 2023, Emmanuel, Drop in gain for DEED 14 alndeed14.pngalndeed14b.pngalndeed14old.png
Dear all,

There was an issue with one of the strips of the deltaE detector (DEED 14), for run216644. The gain dropped to almost a factor of 3. Find attached.

Somehow, it was fixed, but still lags, when compared to other strips of deltaE. This was tested for a more recent run, run216651. Find attached.
Entry  Sun Oct 9 20:57:54 2022, Nikolay Sosnin, Detectors in-beam, no target 12x
After long beam-off, changed detector mount by moving it 2 cm forward into the beam, after removing LiF 3.
Entry  Sat Jul 12 11:34:31 2025, Selin, Detector-Cable Connections Detector-Cable.pdf
List of the detectors and the connections.
Entry  Thu Aug 10 09:20:37 2023, CLW, Detector Signal Cabling Al26 Aug 2023 8x

delta E detector Cabling


Lemo Box Output Number
Signal Cable Bunker
Patch Panel Number
Signal
Cable Rack Area
FADC Number
DAQ Channel




0
72
2204803
9
02911
DEED-1


1
73
2204804
10
02911
DEED-2


2
74
2204805
11
02911
DEED-3


3
75
2204806
12
02911
DEED-4


4
76
2204807
13
02914
DEED-5


5
77
2204808
14
02914
DEED-6


6
80
2204809
15
02914
DEED-7


7
81
2204810
16
02914
DEED-8


8
82
2204811
17
Entry  Thu Aug 10 09:10:02 2023, CLW, Detector Setup Al26 Run 2023 20230809_181749.jpg20230809_181758.jpg20230809_172545.jpg20230809_172442.jpg
dE Detector: W1-20 3186-9 20 um

E Detector: W1-150 3458-4 144 um

Voltages and Leakage Currents: