All AIDA operations are controlled from the Dell T410 Workstation aidas1 using web browsers.
AIDA has multiple, independent data acquisition servers connected by an Ethernet network.
Response times to commands, information updates etc are ~10-60s - please be patient!
How To Start Run
1) Workspace 1 - Firefox web browser - Run Control tab
Select Update
Check DAQ stopped (green GO button displayed)
Check state of *all* data acquisition servers is 'stopped' and histogramming and data transfer
is enabled indicated by 'H X'
See attachment 1
2) Workspace 2 - Firefox web browser - TapeServer tab
Check Reload
Check TapeServer stopped and disk file closed (green GO button displayed)
See attachment 2
3) Workspace 2 - Firefox web browser - MERGE tab
Check Reload
Check Merge Program paused - see banner
Merge State = GOing : paused : xfer enabled : output paused
If not paused select Toggle Merge Program Pause State
See attachment 3
4) Workspace 1 - Firefox web browser - ASIC Control tab
Check Reload
Check 'Act on ALL FEE64 modules?' and wait for request to be completed
Check 'Act on ALL ASICs' and wait for request to be completed
From 'Select required function' select 'Check ASIC Control'
After ~60s a pop-up will appear showing the response from each ASIC of each data acquisition server.
Check all ASICs are OK and status = 0x001F0000.
Dismiss pop-up by selecting OK
See attachments 4-5
5) Workspace 1 - Firefox web browser - Run Control tab
Start DAQ - select green GO button
Wait (~60s) until state of all data acquisition servers (nnaida1, nnaida2 ... etc) updates
Select 'Data Acquisition Statistics'
6) Workspace 1 - Firefox web browser - Statistics tab
Check Reload
Check 'Show ALL Data Acquisition Servers'
From 'Select counter to Monitor' select 'AIDA SYNC'
Select 'Update Once'
Check *all* data acquisition servers have a non-zero rate (typically ~300-500Hz)
See attachment 6
N.B. If one, or more, data acquisition servers have zero rate
return to Run Control tab
stop DAQ (select red STOP button)
check state of *all* data acquisition servers is 'stopped' with histogramming and
data transfer are enabled indicated by 'H X'
start DAQ (select green GO button)
check state of *all* data acquisition servers is 'going' with histogramming and
data transfer are enabled indicated by 'H X'
return to Statistics tab
check AIDA SYNC rates again and if zero rate persists for one, or more, data acquisition
servers call for expert help and do *NOT* proceed!
7) Workspace 2 - Firefox web browser - TapeServer tab
Start TapeServer and open disk file - select green GO button
8) Workspace 2 - Firefox web browser - MERGE tab
Select 'Reload' several times and check all enabled data links (0-23) are active
bright green = active, olive green = inactive, red = disabled
N.B. If some data links are *always* inactive call for expert help and do NOT proceed!
Select Toggle Merge Program Pause State to enable data merging
Select Reload to observe merge rate (data items/s)
Observe active spinners in MERGE and TapeServer server terminal windows adjacent to web browser
See attachment 7
9) Workspace 2 - Firefox web browser - TapeServer tab
Select Reload and observe number of blocks written to disk
10) Workspace 1 - Firefox web browser - Control tab
Select 'Save/Restore Settings'
From 'Save/Restore Module Settings' tab
select 'Reload'
select 'Module Names' whose settings should be saved (usually all)
select 'Generate new DB key' and make a note of the DB key generated
the DB key will have the format YYYYMMMDD-HH.MM.SS
The DB key will be used to name a directory to store AIDA settings
and is created uid: npg, gid: npgstaff, perms 755. However, its
contents are written uid: nfsnobody, gid nfsnobody. The default
directory (DB Key) permissions do not work - change the permissions
to 777 - before attempting to save data to this directory (DB Key).
launch a terminal and change the permissions of the DB key directory using the command
chmod 777 /MIDAS/DB/EXPERIMENTS/AIDA/[YYYYMMMDD-HH.MM.SS]
select 'Save Settings' to save AIDA ASIC, LED and waveform settings to
the directory (DB key)
11) Update EURICA 2016 Elog with the following information
AIDA
run filename (e.g. R99)
[# data blocks]
start date & time
[stop date & time]
ASIC settings DB key
AIDA trigger (e.g. OR DSSSDs #1 - #6)
additional comments
your name
Complete run sheet for written logbook
How to Stop Run
1) Workspace 2 - Firefox web browser - MERGE tab
Pause MERGE input - select Toggle Merge Program Pause State
2) Workspace 1 - Firefox web browser - Run Control tab
Select Update
Stop DAQ - select red STOP button
Wait (~60s) until state of all data acquisition servers (nnaida1, nnaida2 ... etc) updates
Check state of *all* data acquisition servers is 'stopped'
3) Workspace 2 - Firefox web browser - TapeServer tab
Stop TapeServer and close disk file - select red STOP button
4) Complete EURICA 2016 Elog with the following information
AIDA
run filename (e.g. R99)
[# data blocks]
start date & time
[stop date & time]
ASIC settings DB key
AIDA trigger (e.g. OR DSSSDs #1 - #6)
additional comments
your name
Complete run sheet for written logbook
|