Message ID: 260
Entry time: Mon May 30 18:29:54 2016
Reply to this: 261
|
Author: |
AE |
Subject: |
Co/Bi source run (R7) |
|
|
Some comments regarding source run for beta-gamma calibration, to complement EURICA16 elog entry:
http://ribf-exp.riken.jp/elog/EURICA+U-campaign+2016/
This corresponds to run files R7_* (AIDA) and BiCocal_run003.lmd and BiCocal_run004.lmd (run files could
increase for EURICA)
EURICA trigger: AIDA trigger (#1) OR pulser (~25MHz).
Trigger rate: about 170 Hz
AIDA trigger: Detector 1 with low multiplicity condition; N=1, 2 or 3 within window of 5x40ns
AIDA trigger rate: Det1 X = 7500 (nnaida17+nnaida22)
Det1 Y = 800 (nnaida17+nnaida22)
coincidence X+Y= 330
The trigger output of MACB has an inverted logic (0 when trigger, -0.5 V otherwise)
Attached are screenshots documenting conditions during run.
8am: closed file: R7_110.gz in AIDA, BiCocal*005.lmd in EURICA |
|
|
|
|
|
|
|
|
|
|