> Some comments regarding source run for beta-gamma calibration, to complement EURICA16 elog entry:
> http://ribf-exp.riken.jp/elog/EURICA+U-campaign+2016/
>
> This corresponds to run files R7_* (AIDA) and BiCocal_run003.lmd and BiCocal_run004.lmd (run files could
> increase for EURICA)
>
> EURICA trigger: AIDA trigger (#1) OR pulser (~25MHz).
>
> Trigger rate: about 170 Hz
>
> AIDA trigger: Detector 1 with low multiplicity condition; N=1, 2 or 3 within window of 5x40ns
>
> AIDA trigger rate: Det1 X = 7500 (nnaida17+nnaida22)
> Det1 Y = 800 (nnaida17+nnaida22)
> coincidence X+Y= 330
>
> The trigger output of MACB has an inverted logic (0 when trigger, -0.5 V otherwise)
>
> Attached are screenshots documenting conditions during run.
No
The clock is in 10 ns intervals
So 5 => 50ns
The trigger is Fast NIM standard. Talk to Tom
With such a short time ,a lower limit of 1 and high rate discriminators the Trigger will appear true all the time. |