AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
AIDA |
 |
|
Message ID: 305
Entry time: Fri Jun 10 21:07:52 2016
|
Author: |
AE, DK, TD, ML, MA |
Subject: |
Saturday 11 June 2016 |
|
|
04.51 Start run AIDA18. Beam is back from the operators.
Scalers are shown as attachment 1
Leak currents seem okay (all < 5 uA), shown as attachments 2 and 3.
Discrim levels have a bit of fluctuation depending on the module -- attachment 4
nnaida24 has some pauses (attachment 5)
09.35 AIDA20* - AIDA24* test runs with no beam, i.e. noise, betas and low rate pulser
to investigate FEE64 synchronisation
AIDA20* nnaida19 data link to Merge program enabled
AIDA21* - nnaida19 removed from DAQ configuration file (startup.tcl) and server/web browser restarted
12.35 Detector bias & leakage current, FEE temperatures - OK - see attachments 5-7 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|