AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  AIDA  ELOG logo
Message ID: 326     Entry time: Fri Jul 8 18:54:47 2016
Author: TD 
Subject: Analysis July2016/R1 data 
See https://elog.ph.ed.ac.uk/AIDA/324

Analysis of first and last data file - see attachments 1-2
Rates per FEE64 ~20k data items/s

analyser2.f program - see attachment 3

To compile the analyser2.f program (SL6)

gfortran analyser2.f -o analyser2

analyser2 <filename> to produce summary statistics
analyser2 v <filename> to produce verbose analysis

ADC multiplicity as a function of time (1 channel = 0.1ms) for file R1_114 - see attachment 4
Pulser running at ~270Hz clearly identifiable

Conclusion
DAQ ran overnight without stalling
DAQ did not lose synchronisation

Decrease comparator thresholds to increase rates to experiment-realistic levels ~50k data items/s/FEE64
Attachment 1: R1_0.out  3 kB  | Show | Show all
Attachment 2: R1_114.out  3 kB  | Show | Show all
Attachment 3: analyser2.f  17 kB  Uploaded Fri Jul 8 20:27:28 2016  | Show | Show all
Attachment 4: Untitled.png  2.722 MB  Uploaded Fri Jul 8 21:07:15 2016  | Show | Show all
ELOG V3.1.4-unknown