Logged into aidas1 and opened the nnrpi1 relay control window ... powered up the FEE64s.
seperatly connected to ribfdaq and opened a firefox to aidas1:8015.
By the time the firefox had appeared on my desktop the FEE64s had started up.
Ran through the standard startup. Set thresholds to 255 for all 3 sets of comparators. Disabled all discriminators.
Enable histograms.
Started DAQ.
Using ELOG 74 as a guide logged into nnrpi1 and ran two putty sessions ( USB2 and USB3 ).
Based on the instructions in the HV unit manual attached to ELOG75 powered ion all 6 detectors ( current when settled at 100v = 3 to 5uA )
Using spectrum broswer at nnaida6 and ASIC4 to control the slow comparator threshold ..... needed to reach 32 before the pulser became visible.
Using nnaida6:1.3.L found the pulser peak at channel 32136 with a peak width of 67.73 channels.
Then poweroff the HV on all 6 detectors.
Power off the FEE64s.
Logout of all RIKEN machines and consider the best way to take advantage of the situation which allows the system to be operated completely remotely.
|