AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
AIDA |
|
|
Message ID: 348
Entry time: Thu Aug 25 02:10:10 2016
|
Author: |
TD |
Subject: |
Thursday 25 August |
|
|
11.40 R9 continues
ASIC settings 2016Aug20-15.34.53
Discriminator masks -> 0x0
From Correlation Status & Control tab change Trigger delay from 0x180 to 0x5
BNC PB-4 Pulser
Amplitude 90,000
x5 attenuator IN
t_r 50ns tau_d 50us
- polarity
INT ref, delay MIN, tail pulse
External (random) trigger from EG&G Ortec 462 Start output (Range 81.92us) - frequency c. 120Hz
B2F/F11 temp +25.2 deg C, d.p. +14.5 deg C, RH 50.4%
DSSSD bias/leakage current OK - see attachments 1-2
FEE64 temperatures OK - see attachment 3
Some error messages observed for merger - see attachment 4-6
All related to Q5 (nnaida6)
System-wide checks OK - see attachments 7-12
AIDA SYNC, PAUSE, RESUME, Correlation L, SYNC time warp, Good Events, nnaida6, nnaida5 & nnaida20
statistics - see attachments
13-20
Analysis of R9_213 shows data synchronised - see attachment 21
FEE # versus timestamp (0.1ms / channel)
Note nnaida19 not currently configured |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|