Message ID: 383
Entry time: Wed Oct 12 02:58:19 2016
|
Author: |
TD & CG |
Subject: |
Wednesday 12 October |
|
|
10.50 Merger & TapeServer found stopped
connect errors - see attachment 1
DSSSD bias and leakage currents OK - see attachments 2 & 3
Most recent data file R17_440 last accessed 08.22 12.10.16
R17_440 ADC data synchronised
aidas1> ./analyser2 /TapeData/Sept2016/R17_440
*** GREAT format 3.2.0 analyser - TD - May 2014
*** ERROR: READ I/O error: 5002
blocks: 30546
ADC data format: 147630891 ( 1189229.1 Hz)
Other data format: 101593923 ( 818381.9 Hz)
Sample trace data format: 0 ( 0.0 Hz)
Undefined format: 0 ( 0.0 Hz)
Other data format type: PAUSE: 101 ( 0.8 Hz)
RESUME: 101 ( 0.8 Hz)
SYNC100: 1133804 ( 9133.3 Hz)
FEE64 disc: 100456290 ( 809217.8 Hz)
MBS info: 3627 ( 29.2 Hz)
Other info: 0 ( 0.0 Hz)
ADC data range bit set: 151 ( 1.2 Hz)
Timewarps: ADC: 0 ( 0.0 Hz)
PAUSE: 0 ( 0.0 Hz)
RESUME: 0 ( 0.0 Hz)
SYNC100: 0 ( 0.0 Hz)
FEE64 disc: 0 ( 0.0 Hz)
MBS info: 0 ( 0.0 Hz)
Undefined: 0 ( 0.0 Hz)
Sample trace: 0 ( 0.0 Hz)
Timestamp elapsed time: 124.140 s
FEE module #: 1 elapsed dead time 0.728 s
FEE module #: 2 elapsed dead time 0.184 s
FEE module #: 3 elapsed dead time 0.000 s
FEE module #: 4 elapsed dead time 0.000 s
FEE module #: 5 elapsed dead time 1.034 s
FEE module #: 6 elapsed dead time 3.462 s
FEE module #: 7 elapsed dead time 0.000 s
FEE module #: 8 elapsed dead time 0.000 s
FEE module #: 9 elapsed dead time 0.064 s
FEE module #: 10 elapsed dead time 0.003 s
FEE module #: 11 elapsed dead time 0.000 s
FEE module #: 12 elapsed dead time 0.000 s
FEE module #: 13 elapsed dead time 0.000 s
FEE module #: 14 elapsed dead time 0.062 s
FEE module #: 15 elapsed dead time 0.018 s
FEE module #: 16 elapsed dead time 0.005 s
FEE module #: 17 elapsed dead time 0.817 s
FEE module #: 18 elapsed dead time 0.035 s
FEE module #: 19 elapsed dead time 0.048 s
FEE module #: 20 elapsed dead time 0.000 s
FEE module #: 21 elapsed dead time 0.000 s
FEE module #: 22 elapsed dead time 0.103 s
FEE module #: 23 elapsed dead time 0.216 s
FEE module #: 24 elapsed dead time 0.000 s
FEE module #: 25 elapsed dead time 0.000 s
FEE module #: 26 elapsed dead time 0.000 s
FEE module #: 27 elapsed dead time 0.000 s
FEE module #: 28 elapsed dead time 0.000 s
FEE module #: 29 elapsed dead time 0.000 s
FEE module #: 30 elapsed dead time 0.000 s
FEE module #: 31 elapsed dead time 0.000 s
FEE module #: 32 elapsed dead time 0.000 s
*** Program elapsed time: 14.129s ( 2161.951 blocks/s, 135.122 Mb/s)
Conclude DAQ ran without problem for c. 16.5h which is a significant improvement
assuming that it can be consistently reproduced.
11.20 Merger restart did not clear connect/timeout issues
Re-route BNC - BNC RG58 cable with 50MHz clock from BRIKEN to
AIDA Master MACB external clock input (rear panel) so that it will
not be stepped upon.
Power cycle AIDA FEE64s
N.B. I selected 'Switch ALL On' rather than 'Sequence ALL On' by mistake and one
of the USB ac mains controller output 250VAC/16A fuses blew - use 'Sequence ALL On'
TS_SYNC_PHASE 0
FEE64 temperatures OK
*All* system-wide checks OK
B2F/F11 temperature +24.6 deg C, d.p. +8.9 deg C,RH 36.6%
ASIC settings 2016Sep02-18.17.01
slow comparator 0x10 -> 0x0a
LEC/MEC fast comparators # 1, 55 & 58 disabled for all FEE64s
Master MACB serial #13 (see https://elog.ph.ed.ac.uk/AIDA/367)
Master MACB mode 5
14.05 DAQ starts (file Sept2016/R21)
15.30 Sequence of three electrical power dips over 30 minute period. RIKEN public address
system reports that TEPCO (Tokyo Electric Power Company) - the site supplier - is
having region-wide problems.
The third power dip stopped DAQ.
Unable to recover by restarting merger
Power cycle FEE64s
16.20 DAQ restarts (file Sept2016/R22)
19.05 DAQ stops
Because of risk of power outages due to TEPCO distribution problems we have powered
down AIDA and Julabo FL1006 chiller and disconnected from ac mains. |
|
|
|
|
|
|
|
|
|
|