Message ID: 449
Entry time: Thu Nov 10 09:56:20 2016
|
Author: |
TD, OH, ML |
Subject: |
NP1412 R26 R27 |
|
|
start 18.55 10.11.16
stop
AIDA filename R26_0
BRIKEN filename 161109_1852_767778Ni_045
BigRIPS filename 443
ASIC settings 2016Nov07_23.03.06
DSSSD bias & leakage currents OK attachments 1 and 2
Degrader all degraders out
F11 rate 430Hz
BigRIPS setting 767778Ni F5 out, F7 CH2 in, F11 6mm C in
From offline analysis of R26_2
*** DSSSD # 1 count: 459 old count: 225 dt: 1.02 s HEC rate: 229.83 Hz
*** DSSSD # 2 count: 385 old count: 191 dt: 1.02 s HEC rate: 190.54 Hz
*** DSSSD # 3 count: 461 old count: 228 dt: 1.02 s HEC rate: 228.85 Hz
*** DSSSD # 4 count: 403 old count: 197 dt: 1.02 s HEC rate: 202.33 Hz
*** DSSSD # 5 count: 218 old count: 121 dt: 1.02 s HEC rate: 95.27 Hz
*** DSSSD # 6 count: 87 old count: 42 dt: 1.02 s HEC rate: 44.20 Hz
19.02 FEE64 temperature OK - see attachment 3
19.16 BRIKEN DAQ problems following restart - multiple correlation scaler reset requests issued - ignore data
before R26_5
New BRIKEN data file 161109_1852_767778Ni_046
19.20 AIDA DAQ stalled (presumably *end* of LN2 fill cycle)
Soft reset OK
Clock lock state status 0xd - see attachment 6
19.29 Start DAQ (R27)
19.35 BRIKEN DAQ problems starting SyncCheck program - - multiple correlation scaler reset requests issued -
ignore data before R27_2
19.37 All degraders (5.4mm Al) in for new Takechi et al. setting
BigRIPS setting 767778Ni F5 in, F7 out, F11 6mm C in
BRIKEN filename 161109_1852_767778Ni_045
AIDA fileame R27_3
BigRIPS filename 444
19.40 All degraders out
DSSSD bias & leakage currents OK attachments 4 & 5
From offline analysis of R27_4
*** DSSSD # 1 count: 281 old count: 247 dt: 2.31 s HEC rate: 121.43 Hz
*** DSSSD # 2 count: 247 old count: 219 dt: 2.31 s HEC rate: 106.74 Hz
*** DSSSD # 3 count: 286 old count: 252 dt: 2.31 s HEC rate: 123.59 Hz
*** DSSSD # 4 count: 287 old count: 253 dt: 2.31 s HEC rate: 124.02 Hz
*** DSSSD # 5 count: 201 old count: 178 dt: 2.31 s HEC rate: 86.86 Hz
*** DSSSD # 6 count: 33 old count: 29 dt: 2.31 s HEC rate: 14.26 Hz
20.36 Noted x4 increase in F11 rate to c. 2kHz
All degraders in (5.4mm Al)
BRIKEN DAQ shows x4-5 increase for c. 6 minutes
DSSSD bias & leakage currents OK attachments 7 & 8 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|