14.15 BigRIPS run number is now 1029
15.22 BigRIPS run number is now 1030
15.36 DSSSD bias & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
DAQ continues (RIBF03R1/R20)
Degrader 3mm Pb fixed + (0.3mm W/1mm Al + 0.5mm W/1mm Al) variable
ASIC settings 2016Nov17-10.50.10
BNC PB-4 settings
Amplitude 90,000
x5 attenuator IN
freq 25Hz range
t_r 50ns, tau_d 50us
polarity -, int ref, delay MIN, tail pulse
15.43 System-wide checks OK *except* clock status errors from amongst nnaida5, 7, 8, 15 & 16 - bit 2 always set
15.46 Visual scalers - see attachment 4
16.04 Accelerator group wants to change the carbon stripper foil -- beam stop
Run stop (R20_27 at the end)
16.10 Ambient temperature +25.2 deg C, d.p. +2.6 deg C, RH 23.8%
16.17 BigRIPS run 1031 is junk
16:17 Added 1 mm Al plate to the front of the 3 mm Pb degrader (on the upstream side of PE wall)
Pb bricks stacked between MUSIC1 and MUSIC2.
Not enough time to measure it because of quick foil change
Basically, about 10 cm thick, and it has a ~7 cm +/- 1 mm window in X, and no cut in Y (it is above and
below MUSIC window)
16:19 Start AIDA run R21. BRIKEN run started. BRIKEN Sync issued.
16.22 BigRIPS run 1032
16.26 SyncCheck - see attachment 5
17:01 AIDA merger stopped. When attempting to stop the Run, the following error is issued:
STATE for nnaida1 returned with an error
error: SOAP http transport timed out after 20000 ms
NONE
error: SOAP http transport timed out after 20000 ms
while executing
"$transport $procVarName $url $req"
(procedure "::SOAP::invoke" line 18)
invoked from within
"::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetState"
("eval" body line 1)
invoked from within
"eval ::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetState $args"
(procedure "DataAcquisitionControlClient__GetState" line 1)
invoked from within
"DataAcquisitionControlClient__GetState"
Going through hardware checks, the check sync errors since baseline gets 23/24 failures (nnaida6 is still my friend)
Perform TimeStamp RESYNC -> all can pass the check
17:17 AIDA looks happy now. R22 started.
18.16 System-wide checks OK *except* clock status errors from amongst nnaida5, 7, 8, 15 & 16 - bit 2 always set
and ADC calibration fails nnaida13
18.27 DAQ stops OK
Soft reset following LN2 refill cycle
18.33 BigRIPS 1034
18.56 DAQ starts (RIBF03R1/R23)
Degrader (3mm Pb + 1mm Al) fixed + (0.3mm W/1mm Al + 0.5mm W/1mm Al) variable
ASIC settings 2016Nov17-10.50.10
BNC PB-4 settings
Amplitude 90,000
x5 attenuator IN
freq 25Hz range
t_r 50ns, tau_d 50us
19.02 BRIKEN issue correlation scaler reset request
SyncCheck OK - see attachment 6
Visual scaler, 1.8.L & 1.8.W spectra, good events stats - see attachments 7-12
1.8.L pulser peak widths
Peak width = 66.76+/- 0.7221
Peak width = 178.72+/- 2.5140
Peak width = 75.37+/- 0.9037
Peak width = 76.21+/- 0.6630
Peak width = 70.19+/- 0.6772
Peak width = 66.65+/- 0.8516
Peak width = 91.90+/- 0.8132
Peak width = 69.90+/- 1.169
Peak width = 73.73+/- 0.6390
Peak width = 73.01+/- 0.7860
Peak width = 81.35+/- 0.7306
Peak width = 84.03+/- 0.8589
Peak width = 77.27+/- 1.2263
Peak width = 78.92+/- 1.2343
Peak width = 76.99+/- 1.1963
Peak width = 68.85+/- 1.7060
Peak width = 82.37+/- 1.2500
Peak width = 84.83+/- 1.3169
Peak width = 90.84+/- 1.4190
Peak width = 89.81+/- 1.3821
Peak width = 100.33+/- 1.6409
Peak width = 89.42+/- 1.4682
Peak width = 97.52+/- 1.5855
Peak width = 64.02+/- 1.0206
19.19 DAQ stops
LN2 refill transient?
19.28 soft reset
DAQ starts (RIBF03R1/R24)
Degrader (3mm Pb + 1mm Al) fixed + (0.3mm W/1mm Al + 0.5mm W/1mm Al) variable
ASIC settings 2016Nov17-10.50.10
BNC PB-4 settings
Amplitude 90,000
x5 attenuator IN
freq 25Hz range
t_r 50ns, tau_d 50us
19.34 BigRIPS file 1035
19.50 R24_3 AIDA ADC & disc data synchronised - see attachments 13 & 14
21:11 BRIKEN DAQ was not working properly from the last ~half hour (unrelated to AIDA), and had to be restarted and we
also had to send a couple of synchronization pulses.
R25_0: Started new run once BRIKEN was running and all DAQs synchronized
21.41 BigRIPS file 1037
22.14 Stop DAQ to change degrader
22.16 DAQ starts (RIBF03R1/R26)
Degrader (3mm Pb + 1mm Al) fixed + (0.3mm W/1mm Al + 1.0mm W/1mm Al) variable
ASIC settings 2016Nov17-10.50.10
BNC PB-4 settings
Amplitude 90,000
x5 attenuator IN
freq 25Hz range
t_r 50ns, tau_d 50us
22.36 BRIKEN DAQ restarted
re-sync issued
AIDA data file R26_3
BRIKEN data file 039
22.42 BigRIPS file 1038
AIDA data file R26_4 |