10.01 DAQ continues OK
Current file May2017/R8_334
ASIC settings 2017Mar27-16.12.39 *except*
slow comparator 0x0a
LEC/MEC fast comparator 0xff
Detector biases & leakage currents OK - see attachments 1 & 2
FEE64 Temperatures OK - see attachment 3
'good events' statistics OK - see attachment 4
No warnings/errors reported by merger since DAQ GO - see attachment 5
*all* system wide checks OK
10.06 Analysis of data file R8_333 - see attachment 6
dead-time/idle-time < 2%
AIDA SYNC100 variations ~1800/95000
10.10 E17/F11 ambient temperature +24.8 deg C, RH 36.4%, d.p. +8.9 deg C
Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 65%)
10.15 DAQ stops OK
Current file May2017/R7_298
Files May2017/R8* *deleted*
ASIC settings 2017Mar27-16.12.39 *except*
slow comparator *0x08*
LEC/MEC fast comparator 0xff
BNC PB-4 settings
Amplitude 90,000
x5 attenuator IN
t_r 50ns, tau_d 50us
polarity -
rate 250Hz range
delay MIN, tail pulse
Check ASIC Control x2
Zero all spectra
10.20 DAQ starts
File May2017/R8
Transient timeouts at DAQ GO for FEE64s nnaida1 and nnaida2 which resolved themselves without user
intervention
DAQ run control transitioned from 'RESET and STOP/GO' to 'RESET' to 'RESET and STOP/GO'
10.22 DAQ stops
File May2017/R9
Tapeserver data rates fluctuate 10-30Mb/s possibly due to parallel disk I/O activity (deleting R8 files)
Transient timeouts at DAQ STOP for FEE64s nnaida1 and nnaida2 which resolved themselves without user
intervention - see attachment 7
DAQ run control transitioned from 'RESET and STOP/GO' to 'RESET' to 'RESET and STOP/GO'
ASIC settings 2017Mar27-16.12.39 *except*
slow comparator *0x0a*
LEC/MEC fast comparator 0xff
10.31 DAQ starts
File May2017/R10 |