11.19 DAQ continues OK
Current file May2017/R10_346
ASIC settings 2017Mar27-16.12.39 *except*
slow comparator 0x0a
LEC/MEC fast comparator 0xff
Detector biases & leakage currents OK - see attachments 1 & 2
FEE64 Temperatures OK - see attachment 3
'good events' statistics OK - see attachment 4
No warnings/errors reported by merger since DAQ GO - see attachment 5
*all* system wide checks OK
11.22 Analysis of data file R10_246 - see attachment 6
dead-time/idle-time < 1%
AIDA SYNC100 variations ~700/97000
10.10 E17/F11 ambient temperature +24.5 deg C, RH 38.0%, d.p. +9.4 deg C
Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 65%)
11.23 DAQ stops OK
Current file May2017/R10_347
Files May2017/R10* *deleted*
ASIC settings 2017Mar27-16.12.39 *except*
slow comparator 0x0a
LEC/MEC fast comparator 0xff
BNC PB-4 settings
Amplitude 90,000
x5 attenuator IN
t_r 50ns, tau_d 50us
polarity -
rate 250Hz range
delay MIN, tail pulse
Check ASIC Control x2
Zero all spectra
11.30 DAQ starts
File May2017/R11
18.18 Detector biases & leakage currents OK - see attachments 7 & 8
FEE64 Temperatures OK - see attachment 9
18.24 E17/F11 ambient temperature +24.5 deg C, RH 36.8%, d.p. +8.7 deg C
Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 65%) |