11.27 DAQ continues
      Not writing to disk

      BNC PB-4
      Amplitude 90,000
      x5 attenuator IN
      t_r 100ns tau_d 50us
      - polarity
      frequency range 25Hz
      delay MIN, tail pulse, INT ref

      ASIC settings 2017Oct23-14.38.03
      *all* FEE64s slow comparators 0xa

      Detector biases & leakage currents OK - see attachments 1 & 2
      FEE64 Temperatures OK - see attachment 3
      good events statistics stable & OK - see attachments 4
      *all* system wide checks OK *except* nnaida21 fails ADC calibration

11.47 Ambient temperature +23.9 deg C, d.p. +9.0 deg C, RH 39.0%

16:45 System wide checks all ok 
      FEE temps ok - Attachment 5
      Stats - Attachment 6
      Bias and leakage ok - Attachment 7