AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
AIDA |
 |
|
Message ID: 685
Entry time: Fri Oct 27 01:58:44 2017
|
Author: |
TD, OH |
Subject: |
Friday 27 October |
|
|
09.56 DAQ stopped OK yesterday
BNC PB-4
Amplitude 90,000
x5 attenuator IN
t_r 100ns tau_d 50us
- polarity
frequency range 25Hz
delay MIN, tail pulse, INT ref
ASIC settings 2017Oct23-14.38.03
*all* FEE64s slow comparators 0xa *except* nnaida13 0xe, nnaida19-22 0xc
Detector biases & leakage currents OK - see attachments 1 & 2
FEE64 Temperatures OK - see attachment 3
*all* system wide checks OK *except* nnaida21 fails ADC calibration
Julabo FL11006 recirculating chiller coolant reservoir refilled - set point +20 deg C, actual +20.0 deg C
Ambient temperature +23.6 deg C, d.p. +6.2 deg C, RH 32.2%
AIDA withdrawn from (CH2)n matrix
DAQ started - no data transfer
good events statistics stable - see attachment 4
*all* system wide checks OK *except* nnaida21 fails ADC calibration
13:30 System wide checks ok
Stats - Attachment 6
FEE temperatures ok - Attachment 7
leakage and bias ok - Attachment 8
17.15 Detector biases & leakage currents OK - see attachments 9 & 10
FEE64 Temperatures OK - see attachment 11
good events statistics stable - see attachment 12
*all* system wide checks OK *except* nnaida21 fails ADC calibration
Ambient temperature +25.2 deg C, d.p. +5.7 deg C, RH 28.5% |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|