AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
AIDA |
 |
|
Message ID: 697
Entry time: Fri Nov 3 23:09:36 2017
|
Author: |
TD, OH |
Subject: |
Saturday 4 November |
|
|
08.09 DAQ running OK - data transfer not enabled
BNC PB-4
Amplitude 90,000
x5 attenuator IN
t_r 100ns tau_d 50us
- polarity
frequency range 25Hz
delay MIN, tail pulse, INT ref
ASIC settings 2017Oct23-14.38.03
*all* FEE64s slow comparators 0xa *except* nnaida19-22 0xc
Detector biases & leakage currents OK - see attachments 1 & 2
FEE64 Temperatures OK - see attachment 3
DAQ statistics OK - see attachment 4
*all* system wide checks OK *except* nnaida21 fails ADC calibration - see attachment 6
PLL and DCM Clock Lock Monitor OK - see attachment 5
08.22 per FEE64 Rate, 1.8.L and 1.8.W spectra - see attachments 7-12
nnaida3 78 ch FWHM
nnaida17 66 ch FWHM
16.43 Detector biases & leakage currents OK - see attachments 13 & 14
FEE64 Temperatures OK - see attachment 15
DAQ statistics OK - see attachment 16
*all* system wide checks OK *except* nnaida21 fails ADC calibration - see attachment 6
PLL and DCM Clock Lock Monitor OK - see attachment 5
per FEE64 Rate, 1.8.L pectra - see attachments 17-20
nnaida3 78 ch FWHM
nnaida17 68 ch FWHM |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|