AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
AIDA |
 |
|
Message ID: 707
Entry time: Tue Nov 21 00:30:31 2017
|
Author: |
TD, CB |
Subject: |
Tuesday 21 November |
|
|
09.27 Detector biases & leakage currents OK - see attachments 1 & 2
FEE64 Temperatures OK - see attachment 3
DAQ statistics OK - see attachment 4
8x FEE64s <10k
2x FEE64s 10-20k
5x FEE64s 20-30k
3x FEE64s 30-40k
1x FEE64s 40-50k
5x FEE64 >50k (nnaida13 255k)
*all* system wide checks OK *except* nnaida7 fails ADC calibration
09.42 per FEE64 rate, 1.8.L, 1.8.W (20ms FSR) spectra - see attachments 5-10
all spectra zero'd
17.05 Detector biases & leakage currents OK - see attachments 11 & 12
FEE64 Temperatures OK - see attachment 13
DAQ statistics OK - see attachment 14
7x FEE64s <10k
4x FEE64s 10-20k
2x FEE64s 20-30k
4x FEE64s 30-40k
2x FEE64s 40-50k
5x FEE64 >50k (nnaida13 261k)
*all* system wide checks OK *except* nnaida7 fails ADC calibration
per FEE64 rate, 1.8.L, 1.8.W (20ms FSR) spectra - see attachments 15-20
pulser peak width nnaida17 69 ch FWHM, nnaida8 72 ch FWHM |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|