AIDA |
 |
|
Message ID: 722
Entry time: Sun Jun 24 01:40:08 2018
|
Author: |
OH, TD |
Subject: |
Sunday 24 June 2018 |
|
|
09.38 DAQ continues OK
DAQ stop June18/R5_8
Detector biases & leakage currents OK - see attachments 1 & 2
Ambient temperature +25.3 deg C, d.p. +9.4 deg C, RH 36.4%
FEE64 temperatures OK - see attachment 3
System-wide checks OK *except*
nnaida25 failed global clock check code 6
failed ADC calibration
Analysis June18/R5_8 - see attachment 4
10.00 *all* slow comparator thresholds 0x64 -> 0xa
BNC PB-4 pulser ON
per FEE64 1.8.W spectra - 20us & 200us range - see attachments 5-8
14:16 FEE Module nnaida25 & nnaida26 replaced
nnaida25 2b:09:da -> 2b:09;e7
nnaida26 2b:11:c5 -> 2b:22:55
System-wide checks OK *except*
nnaida21 failed global clock check code 6
failed ADC calibration
nnaida28 failed global clock check code 6
failed ADC calibration
511
Since replacing histograming has begun working for all FEE64s and rates across detectors seem lower
Waveforms of FEEs attachments 10 & 11
ADC Rates - attachment 12
Good event statistics - attachment 13
14.28 per FEE64 1.8.W spectra - 200us range - see attachments 14 & 15
17:06 Thresholds at 0x64 across all FEEs
Alpha run started R8 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|