Message ID: 742
Entry time: Sat Sep 29 06:25:23 2018
|
Author: |
OH, TD |
Subject: |
Saturday 29 September |
|
|
14:25 Moved from 8x DSSD setup to 6x DSSD
Top two detectors removed.
Detector alignments in snout unchanged so can use the 6xDSSD mark from June 2018
Rev F adapter cards have been put in place for all backside FEEs
All adapter boards are now Rev E or F
All adapter boards have an additional ground cable from the screw on the copper block to the lemo ground
connection
Power on and detectors biased (Cabling changed to reflect new 1-6)
Leakage and bias ok - attachment 1
System wide checks all ok
FEE temperatures ok - attachment 2
DAQ Started
Stats are looking very good - attachment 3
Rates look ok (Lin and Log) - attachments 4 & 5
Front side waveforms look ok - attachment 6
Backside waveforms look ok - attachment 7
Pulser widths are showing double peaking. - attachments 8 & 9
we are not sure if this is down to the pulser or not
Expanded timescales of the waveform, 0, 10, 100 and 1000 cycles
Layout 7 - attachments 10 - 14
Layout 8 - attachments 15 - 18
15:24 Pulser turned off in prep for alpha run
Slow comp set to 0x64
15.42 DAQ start file Sep18/R2
Test and calibration using background alpha particles
ASIC settings 2018Jun22-11.28.18
*except* slow comparator 0xa -> 0x64
BNC PB-4 pulser OFF
Merger c. 30k items/s TapeServer 230kb/s |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|