AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
AIDA |
 |
|
Message ID: 757
Entry time: Fri Nov 2 00:57:24 2018
|
Author: |
CA, CB, TD, OH |
Subject: |
Friday 2 November |
|
|
09.58 DAQ continues OK - current file Oct18/R15_6
alpha background run
09.20 all system wide checks OK *except*
nnaida7 fails ADC calibration
SYNC errors since baseline - see attachment 1 - baseline not zero'd following reboot of nnaida2 and
setup of all FEE64s yesterday
SYNC errors OK following new baseline check
detector biases & leakage currents OK - see attachments 2 & 3
FEE64 temperatures OK - see attachment 4
good event statistics OK - see attachment 5
nnaida2 rate due to one hot channel
Merger OK - see attachment 6
TapeServer OK - see attachment 7
no errors or warnings shown by server terminal sessions
1.8.L spectra - see attachments 8 & 9
N.B. evidence of minimally ionising particles (presumably protons) from beam to F11 E detector
~10-15cm Pb bricks were placed within BRIKEN (CH2)n stack to stop heavy particles
1.8.W spectra - see attachments 10 & 11 - trace updates for 9/12 p+n FEE64s have stopped, n+n FEE64s OK
15.54 DAQ continues - current file Oct18/R15_15
nnaida17 good event rate c. 400kHz
15.57 FEE64 reset, setup, system wide checks
all ASIC clocks sync'd using ReSYNC
all spectra zero'd
ASIC check load
DAQ go - data transfer disabled
good events statistics - attachment 12
1.8.W spectra - see attachments 13 & 14
changed LED threshold 7700 -> 8500 for 9/12 p+n FEE64s |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|