Message ID: 774
Entry time: Sat Nov 10 21:53:29 2018
|
Author: |
CA, CB, TD, OH |
Subject: |
Sunday 11 November |
|
|
06.53 DAQ continues OK - data transfer disabled
06.55 all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
Julabo FL11006 chiller set point +20 deg C actual +20 deg C, water level c. 75% (L-H)
ambient tempature + 26.9 deg C, RH 21.6%, d.p. +2.9 deg C
FEE64 temperatures OK - see attachment 3
good event statistics OK - see attachment 4
per FEE64 rate spectra - see attachment 5
ASIC check load
per FEE64 rate spectra - see attachment 6
good event statistics OK - see attachment 7
1.8.L spectra - see attachments 8 & 9
nnaida1 pulser peak width 72 ch FWHM
1.8.W spectra 20us FSR - see attachments 10 & 14
1.8.W spectra 200us FSR - see attachments 11 & 15
1.8.W spectra 2ms FSR - see attachments 12 & 16
1.8.W spectra 20ms FSR - see attachments 13 & 17
Note - LED threshold 9000 for *all* FEE64s
07.27 statistics zero'd
some 32bit counters (e.g. nnaida19 & nnaida20) had overflowed/wrapped around
18:19 Computer appears to be running slow though cannot see a reason why.
No programs using to much ram or cpu
FEE temp ok - attachment 18
bias and leakage currents ok - attachment 19
System wide checks ok
21:02 System wide checks ok
DAQ Continues ok - data transfer disabled.
Bias and leakage currents ok - attachment 20
FEE Temperatures ok - attachment 21 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|