11.00 - chiller refilled and switched on
set to 20 degrees celsius
AIDA stand moved and secured in F11
clock connected -> 50 MHz square wave at 5.0 V
AIDA connected to power box
network cables attached
NIM modules switched on
Raspberry Pis switched on
Detector bias switched on for detectors 1-6
13.58 - AIDA power relay 1,2,5,6 sequenced on
bias/leakage currents ok (attachment 1)
DAQ setup
System wide checks ok, except naida07 fails ADC calibration, and clock status check returns error in
nnaida06, 5.
Above issue with clock no longer present
FEE temperatures ok (attachment 2)
good event statistics (attachment 3)
16.30 1.8.W spectra n+n strips
FSR 20us, 200us, 2ms, 20ms attachments 4-7
1.8.W spectra p+n strips
FSR 20us, 200us, 2ms, 20ms attachments 8-11
compare with https://elog.ph.ed.ac.uk/AIDA/746
naida19, naida20, naida23 show low frequency noise structure, worse than in above elog entry
17.22 1.8.L spectra p+n strips
peaks:
aida01 72
aida02 112
aida03 80
aida04 78
aida05 74
aida06 95
aida07 78
aida08 71
aida09 82
aida10 85
aida11 93
aida12 96
aida13 101
aida14 70
aida15 87
aida16 87
aida17 69
aida18 83
aida19 127
aida20 116
aida21 84
aida22 74
aida23 108
aida24 117
|