AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
AIDA |
 |
|
Message ID: 800
Entry time: Wed May 15 02:17:23 2019
|
Author: |
CA, OH, TD |
Subject: |
Wednesday 15 May |
|
|
10.18 DAQ continues OK to file May19/R2
alpha background
ASIC settings 2018Oct03-13.26.58
*except* slow comparator 0xa -> 0x64
BNC PB-4 Pulser OFF
all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
good event statistics OK - see attachments 4
Merger OK - see attachment 5
TapeServer OK - see attachment 6
1.8.L spectra - attachments 7 & 8
1.8.W spectra 20us FSR - attachments 10 & 11
1.8.W spectra 20ms FSR - attachments 10 & 11
Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C - water topped up
Temperature + deg 24.7C dew point +7.5 deg C RH 33.3%
FEE64 nnaida1-4 temperature logs - see attachments 12-15
all FEE64 temperatures appear to be stable
11.05 check ASIC control *all* FEE64s and ASICs
11.13 analysis of file May19/R2_1 - see attachment 16
ADC rate c. 72Hz dominated by same, single hot channel (Stat spectrum ch# 104) in each of FEE64s nnaida1,
2, 5 & 6 otherwise looks OK
sync pulser rate 10Hz confirmed by JA
17.25 all system wide checks OK *except* nnaida7 fails ADC calibration
FEE64 temperatures OK - see attachment 18
good event statistics OK - see attachments 19
detector biases & leakage currents OK - see attachment 20
Temperature + deg 24.5C dew point +8.2 deg C RH 34.1%
Merger and TapeServer ok |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|