Message ID: 801
Entry time: Thu May 16 02:13:10 2019
|
Author: |
CA, OH, TD |
Subject: |
Thursday 16th May |
|
|
10:13 all system wide checks OK *except* nnaida7 fails ADC calibration
FEE64 temperatures OK - see attachment 1
good event statistics OK - see attachments 2
detector biases & leakage currents OK - see attachment 3
Temperature + deg 24.7C dew point +8.6 deg C RH 35.0%
Merger and TapeServer ok
10:30 DAQ, merger, TapeServer stop
Alpha run stopped at AIDA file TapeData/May19/R2_5
slow comparator threshold lowered to 10 keV (0xa)
DAQ start, not writing to file
good event statistics after lowering threshold, ok (attachment 4)
10:49 1.8.W spectra 20ms FSR - attachments 5 & 6
11:25 started writing to file R3 (sync test)
low threshold run at 5mb/s
11:33 DAQ stop (R3)
slow comparator threshold raised to 100 kev (0x64)
DAQ start, alpha run
Writing to file TapeData/May19/R4
18.49 alpha background
ASIC settings 2018Oct03-13.26.58
*except* slow comparator 0xa -> 0x64
BNC PB-4 Pulser OFF
FEE64 temperatures OK - see attachment 7
good event statistics - see attachments 8 - nnaida1 405k & nnaida17 368k, all other FEE64s c. 380Hz as
expected
ASIC check for *all* ASICs & FEE64s restored all FEE64 rates to expected c. 380Hz - see attachment 13
detector biases & leakage currents OK - see attachments 9 & 10
Merger OK - no error messages - see attachment 11
TapeServer OK - see attachment 12
all system wide checks OK *except* nnaida7 fails ADC calibration *and* Check Sync errors not counted Since
Baseline - see attachment 14
DAQ stopped May19/R4_3
aidas1> ls -l /TapeData/May19
total 19808048
-rw-rw-rw-. 1 npg npgstaff 1908736000 May 14 18:28 R1_0
-rw-rw-rw-. 1 npg npgstaff 2097152000 May 15 02:30 R2_0
-rw-rw-rw-. 1 npg npgstaff 2097152000 May 15 10:20 R2_1
-rw-rw-rw-. 1 npg npgstaff 2097152000 May 15 18:08 R2_2
-rw-rw-rw-. 1 npg npgstaff 2097152000 May 16 01:57 R2_3
-rw-rw-rw-. 1 npg npgstaff 2097152000 May 16 09:45 R2_4
-rw-rw-rw-. 1 npg npgstaff 190447616 May 16 10:28 R2_5
-rw-rw-rw-. 1 npg npgstaff 1109327872 May 16 11:28 R3_0
-rw-rw-rw-. 1 npg npgstaff 2097152000 May 16 18:39 R4_0
-rw-rw-rw-. 1 npg npgstaff 2097152000 May 16 18:45 R4_1
-rw-rw-rw-. 1 npg npgstaff 2097152000 May 16 18:50 R4_2
-rw-rw-rw-. 1 npg npgstaff 225312768 May 16 18:54 R4_3
-rw-rw-rw-. 1 npg npgstaff 52494336 May 16 19:10 R5_0
Wahtever the issue was it probably occurred before 18.39
DAQ RESET
DAQ SETUP
ASIC check for *all* ASICs & FEE64s
all system wide checks OK *except* nnaida7 fails ADC calibration
Sync all ASIC clocks using Re-SYNC
DAQ started May19/R5
19.23 1.8.W spectra - see attachments 15 & 16 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|