Message ID: 807
Entry time: Mon May 20 02:16:58 2019
|
Author: |
CA, TD |
Subject: |
Monday 20 May 2019 |
|
|
10:18 DAQ Continues ok
Alpha run continues - Currently on file R8_2
System wide checks all ok
Detector bias/leakage currents ok - attachment 1
Temp +25.6 deg C, Humidity 34.5% AND D.P +8.8 deg C
FEE Temperatures ok - attachment 2
Good event statistics ok - attachment 3
Merger & TapeServer ok - attachments 4 & 5
17.15 DAQ continues OK to file May19/R8_2
alpha background
ASIC settings 2019May19-17.54.03
slow comparator 0x64
BNC PB-4 Pulser OFF
all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 6 & 7
FEE64 temperatures OK - see attachment 8
good event statistics OK - see attachments 9
Merger OK - see attachment 10
TapeServer OK - see attachment 11
1.8.L spectra - see attachments 12 & 13
p+n FEE64s 1.8.W spectra 20us FSR - attachment 14
n+n FEE64s 1.8.W spectra 20us FSR - attachment 15
17.25 analysis of May19/R8_2 - see attachment 16
deadtime OK
ADC data for FEE64s nnaida5 & nnaida6 hot channels (Stat spectrum channel # 104)
Also observed at a much lower rate for nnaida1 & nnaida2
17.27 Check ASIC Control *all* ASICs & FEE64s
13.27 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
Water level c. 60% M-H markers - water topped up (1x jugful) to H level marker
Temperature +25.4 deg C dew point +9.6 deg C RH 36.3% |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|