Monday 28 May |
|
|
09.30 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
Water level c. 75% M-H markers
Temperature +25.1 deg C dew point +9.7 deg C RH 37.9%
DAQ continues OK
all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
good event statistics OK - see attachments 4
13x < 20k, 22x <50k, 24x <100k
1.8.L spectra - see attachments 5 & 6
nnaida1 73 ch FWHM, nnaida3 78 ch FWHM
p+n FEE64s 1.8.W spectra 20us FSR - attachment 7
n+n FEE64s 1.8.W spectra 20us FSR - attachment 8 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|