14.10 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
Water level c. 75% M-H markers
Temperature +25.7 deg C dew point +8.2 deg C RH 32.4%
DAQ continues OK
ASIC settings 2019May19-17.54.03
slow comparator 0xa
BNC PB-4 Pulser
Amplitude 90,000
x5 attenuator IN
Frequency 25Hz range
t_r 100ns, tau_d 50us
- polarity
Delay MIN, tail pulse
14.14 all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
good event statistics OK - see attachments 4
11x <20k, 22x <50k, 24x <100k - 13x <20k yesterday
1.8.L spectra - see attachments 5 & 6
nnaida1 73 ch FWHM, nnaida3 78 ch FWHM - no change cf. yesterday
p+n FEE64s 1.8.W spectra 20us FSR - attachment 7
n+n FEE64s 1.8.W spectra 20us FSR - attachment 8
FEE64 Rate spectra - attachment 9
Check ASIC Control *all* FEE64s & ASICs
zero all spectra
16.05 slow comparator 0x8
good event statistics OK - see attachments 10
2x <20k, 11x <50k, 18x <100k
slow comparator 0x9
good event statistics OK - see attachments 11
6x <20k, 19x <50k, 23x <100k
slow comparator 0xa
good event statistics OK - see attachments 12
10x <20k, 21x <50k, 24x <100k |