Message ID: 886     Entry time: Wed Aug 9 14:40:28 2023
Author: Carole, Nic, Jeroen 
Subject: AIDA HV SUPPLY & PULSER TEST 

Tests with the HV supply (CAEN N1419ET) and Pulse Generator ( BNC PB-5 ) in the detector lab on the crate behind the DEGAS detectors, carried out the 9th August 2023

Both modules were plugged

Test 1: pulse generetor on 2 Hz

- DC, 50 ohm - HV OFF - attachment 1

- DC, 50 ohm - HV OFF, Zoom on the noise - attachment 2

- DC, 50 ohm - HV 50 V, Zoom on the noise - attachment 3

 

Test 2: HV supply Channel 1 - CORE and 2 - BRAID, pulse generator still on

- AC 1 Mohm - HV OFF - attachment 4

- AC 1 Mohm - HV 50 V - attachment 5

No differences were observed between crate OFF, HV OFF and HV 50 V. An important periodic noise (green on the pictures) was observed and was significantly smaller when unpluging the HV module from the crate. The thickness of the baseline noise seems to be comparable to entry 882, attachment 33-35.

 

Attachment 1: AIDA_PulserTest_0908_FullPulse.png  84 kB  Uploaded Wed Aug 9 15:42:11 2023  | Hide | Hide all
AIDA_PulserTest_0908_FullPulse.png
Attachment 2: AIDA_PulserTest_09.08_5mV_10us_HVOFF.png  93 kB  Uploaded Wed Aug 9 15:42:56 2023  | Hide | Hide all
AIDA_PulserTest_09.08_5mV_10us_HVOFF.png
Attachment 3: AIDA_PulserTest_0908_10mV_10us_50HV.png  98 kB  Uploaded Wed Aug 9 15:43:07 2023  | Hide | Hide all
AIDA_PulserTest_0908_10mV_10us_50HV.png
Attachment 4: HV_0V_090823_1500_wPulser.png  142 kB  Uploaded Wed Aug 9 15:43:20 2023  | Hide | Hide all
HV_0V_090823_1500_wPulser.png
Attachment 5: HV_-50V_090823_1500_wPulser.png  120 kB  Uploaded Wed Aug 9 15:43:30 2023  | Hide | Hide all
HV_-50V_090823_1500_wPulser.png
ELOG V3.1.4-unknown