AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
AIDA |
 |
|
Message ID: 892
Entry time: Thu Sep 28 16:03:40 2023
|
Author: |
TD |
Subject: |
Thursday 28 September |
|
|
17.00 USB-controlled ac mains relay - no water interlock ( yellow ) LED
Water temperature ( station outside S4 entrance ) c. 19 deg C and pressure OK
Flow can be felt on rubber outflow pipe from water manifold
Failure of sensor or interlock box?
*** Bypass interlock
16.58 Power ON
17.05 FEE64 temperatures OK - see attachment
All system wide checks OK
per FEE64 p+n 1.8.W spectra - attachments 2-42021Apr29-13-16-00
per FEE64 p+n 1,8.L spectra - attachment 5
aida01 pulser peak width 15 ch FWHM
per FEE64 Rate spectra - attachment 6
BNC PB-5 pulser
Amplitude 10.0V
Attenuation x10
Decay 1ms
Frequency 25Hz
Polarity + ( n+n Ohmic test input from EG&G Ortec 534 Sum & Invert Amplifier not connected )
Clamp ON
ASIC settings 2021Apr29-13-16-00
slow comparator 0x64
17.28 FEE64 temperatures - attachment 7
all < 1 deg C difference cf. 17.05
ADC data item stats OK - attachment 8
19.06 FEE64 temperatures OK - attachment 9 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|