AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
AIDA |
 |
|
Message ID: 893
Entry time: Sat Sep 30 12:14:52 2023
|
Author: |
TD |
Subject: |
Satuurday 30 September |
|
|
12.54 Power up
All system wide checks OK
per FEE64 p+n 1.8.W spectra - attachments 5-10
20us, 200us and 2000us FSR
per FEE64 p+n 1,8.L spectra - attachment 3
aida01 pulser peak width 15 ch FWHM
per FEE64 Rate spectra - attachment 4
BNC PB-5 pulser
Amplitude 10.0V
Attenuation x10
Decay 1ms
Frequency 25Hz
Polarity + ( n+n Ohmic test input from EG&G Ortec 534 Sum & Invert Amplifier not connected )
Clamp ON
ASIC settings 2021Apr29-13-16-00
slow comparator 0x64
FEE64 temperatures OK - see attachment 2
ADC data item stats OK - see attachment 1
MSL type BB18 24cm c 8cm DSSSD 3208-3/-21/-22
p+n junction side - 21 Ohm resistnace between bias pins
n+n Ohmic side - 152 Ohm resistance between bias pins
=> all bias bond wires OK |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|