Message ID: 903
Entry time: Tue Oct 10 08:38:37 2023
|
Author: |
TD |
Subject: |
Tuesday 10 October |
|
|
09.38 Water flow & temp OK
Power ON
All system wide checks OK
DSSSD HV +120V +5.585uA - attachment 1
CAEN N1419ET grounded output ( internal LK fitted ) plus ThorLabs inline filter
standard coaxial cable to aida04, no HV daisy chain aida12, aida03 & aida11
aida03 LK3 fitted
aida11, aida03, aida12 LK1 fitted
test + & test - daisy chains removed
FEE64 temps OK - attachment 2
slow comparator 0x64
ADC data item stats - attachment 3
per FEE64 Rate spectra - attachments 4
1.8.W spectra - 200us FSR - attachments 5-8
test + to aida01 *only*
ADC data item stats - attachment 9
per FEE64 Rate spectra - attachment 10
aida01 1*L spectra - attachment 11
pulser peak width 1.8.L 141 ch FWHM
ThorLabs inline filter removed
ADC data item stats - attachment 12
per FEE64 Rate spectra - attachment 13
pulser peak width 1.8.L 136 ch FWHM
14.15 DSSSD HV +120V +6.610uA
Disconnect aida01 ribbon cable drain wire from adaptor PCB gnd
aida01 1.8.L pulser peak width 188 ch FWHM - amplitude/structure of noise does not change significantly but observe higher frequency components in addition
1.8.W p+n spectra - 20us FSR - attachment 14
15.20 Switch from HV + to HV - N1419ET ch 3 to ch 2
HV coaxial cable chain aida12, aida03, aida11, aida04 LK1 fitted
aida12, aida03 & aida11 LK1 removed
Connect aida01 ribbon cable drain wire to adaptor PCB gnd
DSSSD HV-120V -6.720uA - attachment 15
ADC data item stats - attachment 16
aida01 1.8.L pulser peak width 138 ch FWHM
1.8.W spectra - 20us FSR - attachments 17 & 18
Little observable difference between HV - and +, as expected
16.20 Jumper aida04 adaptor PCB gnd to aida12 adaptor PCB gnd
slow comparator 0x64 -> 0xf
ADC data item stats - attachment 20
per FEE64 Rate spectra - attachment 21 & 24
1.8.W spectra - 20us FSR - attachments 22 & 23
aida01 1.8.L pulser peak width 99 ch FWHM
17.00 Ground cable from aida04 gnd to aida12 gnd
ADC data item stats - attachment 25
per FEE64 Rate spectra - attachment 26 & 27
1.8.W spectra - 20us FSR - attachments 28 & 29
aida01 1.8.L pulser peak width 97 ch FWHM
17.20 Ground cable from aida04 gnd to aida12 gnd to aida02 gnd
slow comparator 0xf ~ 150keV
ADC data item stats - attachment 30
per FEE64 Rate spectra - attachment 31
1.8.W spectra - 20us FSR - attachments 32 & 33
1.8.W spectra - 200us FSR - attachments 32 & 33
aida01 1.8.L pulser peak width 85 ch FWHM ~ 65keV FWHM => 5s threshold 138keV
aida12 1.8.L pulser peak width 58 ch FWHM ~ 44keV FWHM => 5s threshold 93keV
aida02 1.8.L pulser peak width 133 ch FWHM ~ 100keV FWHM => 5s threshold 212keV
Rate spectra are consistent with these estimates, i.e. observe ~zero rate for p+n channels ( except hot channels ) and low ( ~100-200Hz rate per channel ) for n+n channels
p+n noise aida12, aida03, aida11 excellent
for aida10, aida01 & aida09 good - can probably be improved to match aida12 etc
significant improvement observed for n+n side
aida12 *.*.W spectra - attachments 34-37
per p+n FEE64 1.8.W spectra - 200us FSR - expanded scale - attachment 38 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|