AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
AIDA |
 |
|
Message ID: 924
Entry time: Sat Nov 18 02:28:26 2023
|
Author: |
TD, NH, JB |
Subject: |
Saturday 18 November |
|
|
03.20 c. 3k per spill 13N beam to S4
Observe scatter of events at all energies for LEC
Switch LE -> MEC
aida02 & aida04 slow comparator 0xf -> 0x3 ( c.1.5MeV )
All other ( p+n ) FEE64s slow comparator 0xa ( c. 5MeV )
For c. 700MeV/u 14N + 6cm s.steel -> c. 350MeV/u 14N -> c. 30MeV energy loss in 1mm Silicon, c. 45MeV energy loss in 3mm plastic
per n+n FEE64 1.8L spectra - attachment 1
observe 3x peaks above threshold - different Z fragments?
per FEE64 Stat spectra
04.10 DSSSD bias -150V leakage current -5.380uA - attachment 3
FEE64 temps OK - attachment 4
ADC data item stats - attachment 5
per FEE64 Stat spectra - attachment 6
per FEE64 1.8.L spectra - attachments 7-8
per FEE64 1.8.W spectra - 20us FSR - attachments 9-10
05.03 aida02 & aida04 slow comparators 0x3 -> 0xa
All spectra zero'd
c. 06.30 Beam tests end
DAQ stopped. DSSSD HV OFF. AIDA power OFF.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|