AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  BRIKEN  ELOG logo
Message ID: 207     Entry time: Wed Mar 29 03:02:02 2017
Author: J. Agramunt 
Subject: Synchronization test procedure (update)  
To Launch the processes:

Open 4 terminals:

Terminal 1 (DataSink) type:

>ssh aida@d05

>cd /home/aida/DataPackage/DataSink/Linux64

>DataSink -i 3

Terminal 2 (SynCheck) type:

>ssh aida@d05

>cd /home/aida/SyncCheck

>./SynCheck

Terminal 3 (BigRIPS Data Relay) type:

>ssh aida@d05

>cd /home/aida/ribfts/ribftssender

>./ribftssender_mar17ca debug

(specific for current 40Ca parasitic run because we are not sending
a pulse signal to  BigRIPS as trigger but receiving triggers from
BigRIPS to create the pulses)

Terminal 4 (AIDA Data Relay) type:

>npg@aidas1
>DataRelayFilter -n 10.32.0.12 -p 10307 -I 2

"-I 2" selects ID 2 as the one for the AIDA data stream, 
which must correspond to the one expected by the DataSink in the d05 PC.
The programming transmitting the data from AIDA to the DataSink includes an 
intermediate filtering step. This selects only correlation scalers from one 
module (and possibly ADC values from one channel in that module).

Gasific MainWindow (Briken Data Relay):

   In the DAQ Control tab: Click "Start Sync Monitoring" button.

   In the pop up window set server ip and port (ip:10.32.0.12, port:10305)

By default this perform a check on the differences RIBF-BRIKEN and AIDA-BRIKEN.

In case that BRIKEN is not running, the option "SynCheck -M 1" check the differences 
AIDA-RIBF and BRIKEN-RIBF, while "SynCheck -M 2" checks RIBF-AIDA and BRIKEN-AIDA

To synchronize:

   In the Gasific MainWindow, DAQ Control tab: Click "Send Sync Pulse" button
ELOG V3.1.4-unknown