Eff callibration parameters for D4-Clover |
|
|
Efficiency calibration for the D4 clover (data acquired in October).
The attached plots show the parameters for the Jackel function (E[i]=Energy) using the points of 152Eu, 60Co and 137Cs sources.
Roger
> Efficiency calibration
>
> File "cloversEff.ods" and "cloversEff.xls" contains the experimental efficiency from different measurements
> during July and October
> (G7 broke before Eu-152 measurement). We can assume the efficiency didn't change from July measurements to
> October measurements. |
|
|
|
|
|
|
|
|