Mon May 29 05:40:49 2017, K. Matsui, energy loss at logn
|
|
Mon May 29 04:17:34 2017, YS, Clover Efficiency  
|
Clover gamma detection efficiency calculation with 152Eu
Date of the measurement: 2017-05-26 10:15
root file: 170526_1015_152Eu_noAIDA_z0cm_014.root
|
Sun May 28 11:41:24 2017, YS, CLOVER energy calibration
|
Clover calibration with 152Eu
Used peaks [keV]: 121.7817, 244.6975, 344.2785, 444.000, 778.9045, 867.378, 964.100, 1112.074, 1408.011
|
Sat May 27 11:19:57 2017, AIM, AT, AIDA+BRIKEN background
|
Background measurement AIDA+BRIKEN
AIDA is inside BRIKEN
|
Sat May 27 10:36:20 2017, JA,AT,JLT, BRIKEN DACQ 
|
Manuals for BRIKEN DACQ:
GasificShort.pdf including "Run protocol"
and Gasific70Manual.pdf (long) |
Sat May 27 04:46:55 2017, AIM, sources info
|
Here is the sunflower webpage with information on the calibration sources:
http://www.nishina.riken.jp/collaboration/SUNFLOWER/misc/util/ribfsource.html |
Fri May 26 14:42:12 2017, JL, AT, Background
|
Background measurement (no AIDA)
|
Fri May 26 14:27:08 2017, AIM, AT, BRIKEN Configuration File
|
New configuration file of BRIKEN
Changes of 26th-May-2017
Previous config file 24th-May-2017
|
Fri May 26 09:07:46 2017, J. Agramunt, Synchronization Monitoring 
|
We started the synchronization monitoring programs to check that
it is OK as can be seen in the figure below
|
Fri May 26 04:20:54 2017, JLT,AA,AT, CLOVER 60Co
|
60Co No AIDA, Efficiency calibration
Start: 12:21
|
Fri May 26 03:52:23 2017, AT,AA,JLT, CLOVER, 137Cs
|
137Cs, No AIDA, Efficiency calibration
Source ID: 393
|
Fri May 26 03:24:38 2017, GK,VP, AT, AA, CLOVER, 133Ba
|
133Ba source efficiency calibration, no AIDA
Start: 11:23
|
Fri May 26 02:17:40 2017, AT, AIM, JLT,AA, PV, KM, 152Eu
|
Calibration with 152Eu source.
START time:10:15
|
Thu May 25 14:31:46 2017, AT, JLT, AA, AIM, Background measurement 7x
|
Background, No AIDA
Start:22:31
|
Thu May 25 13:56:00 2017, AT, JLT, AA, 252Cf
|
252Cf source in Z=0cm position.
Start:21:57
|
Thu May 25 13:42:45 2017, AT, PC,KR,JLT,AA, 252Cf
|
252Cf source in Z=0cm position.
Start:21:41
|
Thu May 25 13:38:11 2017, AT, PC, KR, JLT, AA, Clover, detectors in new position  
|
Both clover detectors were moved. They are in a symmetrical position. The end on the clover nose is of the order
of 3-4 mm from the AIDA tunnel.
|
Thu May 25 13:24:02 2017, AT,PC,KR,JLT,AA, CLOVER, detector not in its rigth position
|
We have noticed that D4 is not in its right position. That explains the different number of counts in the
detectors.
Probably of the order of 3 cm away from its nominal place.
|
Thu May 25 12:53:47 2017, AT, PC, JLT, KR, AA, Clover, 60Co
|
60Co source
Source ID: CD395
|
Thu May 25 12:49:55 2017, AT, JLT,PC,AA, Clover, 133Ba, LN2 filling
|
We started the measurement of the source when the N2 filling started. Just to check if there are funny things
related to the filling.
|
|