> Initial grounding see final grounding from elog 187. ASIC settings for all - LEC slow comparator 0x64. Note on 11/02/22 1.8*L widths with same settings were aida1-4: 300,240,250,260
ADC data items <10k
Good events fluctuate 10k-20k
1.8*L pulser widths aida1-4: 433,400,375,375(double peak)
Attachments 1-4
> New grounding implemented. 3mm grounding cable connect fee body to adaptor board. Each adaptor board is then connected directly to the cryring ground. Attachements 10-11
ADC data items <10k
Good events fluctuate 10k-20k
1.8*L pulser widths aida1-4: 340,288,288,295 (reduced double peak)
Noise in the system also seems to have decreaed in zoomed out 1.8*L plots (not attached). Aida1 still shows large noise peak channel 32000 on 1.8*L. 1.8*W no significant change, perhaps slightly reduced.
Attachements 5-9
Screencaps for following movement tests not taken as 1.8*L and 1.8*W spectra appeared the very similar as 5-9 except for slightly different widths.
> Detectors moved in and out (pneumatic). Leakage current dropped by ~0.2 uA, torque now reads +2.8 previously was ~0.
no change to ADC data items or good events. During movement aida3 saw momentary increase in good events to ~40k before returning to normal range. Rest of fees good events stats unaffected by movement.
1.8*L pulser widths aida1-4: 413,313,295,306
> servo motor (only) moved, torque reduced back to ~0
no change in good events stats for any fees during servo movement.
1.8*L pulser widths aida1-4: 425,335,316,288
> Detector moved to IN position (pnuematic and servo). 19.3mm to 81.8mm on potentiometer.
No change to good events. During movements aida3 saw momentary increase in good events to ~41k, all other fees unaffected.
1.8*L pulser widths aida1-4: 302,221,305,293
Noise peak at channel ~32000 on aida 1 (1.8*L) significantly reduced. little noise on other fees
>Detector moved to fully OUT position (pneumatic and servo).
no change in good events, during movement aida3 saw momentary increase to ~35k
1.8*L pulser widths aida1-4: 450,350,316,290
Noise peak at chn 32000 on aida 1 has increased to previous size. little noise on other fees.
> Bias increased to 120V. Leakage current stable however slightly higher ~0.1 uA
no change in good events.
1.8*L pulser widths aida1-4: 447,345,189,301
> Bias increased to 140V, increase in leakage current ~0.2 uA, appears stable. Increase to 150V, fluctuate in leakage current of +/-0.5uA then huge increase to >20uA trip.
> Fee power cycle. Bias 100V, leakage current 9.9 uA ~0.1-0.2 uA higher than previous but stable.
no change to good events
1.8*L pulser widths aida1-4: 468, 350,311, 296
> Some decrease seen with new grounding, however main source of noise still apparent and any decrease seems insignificant compared to changes due to position movement.
> Water pipes have been zipped tied so they no loger touch motors, no change in pulser widths or noise levels however. |