Message ID: 364     Entry time: Wed Mar 2 14:15:16 2022
Author: JM, PM 
Subject: Current status 

15:15

Temperature setpoint = 113 K

Target density is OK for now. S1 vaies between 2E-8 and 3.5E-8 mbar. E4 between 1E-9 and 1E-8 mbar.

Detector motors functioning well. Moving between 23.1 mm and 67.43mm. Move in and move out counters the same. Correlation scalar 8 for detectors moving is seen on aida1

Rutherford rate peak varies between 10^3 peak height when fully in and ~10^2 when fully out. No peak from halo seen in rate.

Beam intensity ~ 1E8

Merger OK

Data rate ~ 1700 kb/sec

Run is Run 87 - started 10:34

Data spy restarted at 15:00 so only include Run87 data.

Leakage current = 3.205 uA

Fee Temperatures OK

System checks OK - wr fails

         Base         Current     Difference
aida01 fault      0xab60 :      0xaba3 :      67  
aida02 fault      0x88d3 :      0x8914 :      65  
aida03 fault      0x23aa :      0x23ed :      67  
aida04 fault      0x15d7 :      0x161a :      67  
aida05 fault      0x2f03 :      0x2f20 :      29  
White Rabbit error counter test result: Passed 0, Failed 5

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

Avg ring Pressure = 5.5E-10 mbar

16:45

Temperature setpoint = 113 K

Target density is OK for now. S1 vaies between 2E-8 and 3E-8 mbar. E4 between 1E-9 and 1E-8 mbar.

Detector motors functioning well. Moving between 23.1 mm and 67.43mm. Move in and move out counters the same. Correlation scalar 8 for detectors moving is seen on aida1

Rutherford rate peak varies between 10^3 peak height when fully in and ~10^2 when fully out. No peak from halo seen in rate.

Beam intensity ~ 1E8

Merger OK

Data rate ~ 1300 kb/sec

Temperatures OK

Leakage current ~3.205 uA

System chacks OK - wr fail

         Base         Current     Difference
aida01 fault      0xab60 :      0xaba3 :      67  
aida02 fault      0x88d3 :      0x8914 :      65  
aida03 fault      0x23aa :      0x23ed :      67  
aida04 fault      0x15d7 :      0x161a :      67  
aida05 fault      0x2f03 :      0x2f20 :      29  
White Rabbit error counter test result: Passed 0, Failed 5

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

Avg ring pressure ~ 4.4 E-10 mbar

Statistics similar to previous check.

 

Attachment 1: 2022-03-02_15-23-02-ExRate.png  160 kB  | Hide | Hide all
2022-03-02_15-23-02-ExRate.png
Attachment 2: 2022-03-02_15-24-14-XYtotal.png  311 kB  | Hide | Hide all
2022-03-02_15-24-14-XYtotal.png
Attachment 3: 2022-03-02_15-28-04-correlation8.png  103 kB  | Hide | Hide all
2022-03-02_15-28-04-correlation8.png
Attachment 4: 2022-03-02_15-27-13-ADCitems.png  103 kB  | Hide | Hide all
2022-03-02_15-27-13-ADCitems.png
Attachment 5: 2022-03-02_15-26-34-GoodEvents.png  104 kB  | Hide | Hide all
2022-03-02_15-26-34-GoodEvents.png
Attachment 6: 2022-03-02_16-51-19-aida1-4-stats.png  171 kB  Uploaded Wed Mar 2 15:51:32 2022  | Hide | Hide all
2022-03-02_16-51-19-aida1-4-stats.png
ELOG V3.1.4-unknown