CARME |
 |
|
Message ID: 370
Entry time: Tue Aug 9 13:44:16 2022
|
Author: |
TD, JM |
Subject: |
Tuesday 9 August |
|
|
14.00 CARME vacuum c. 1.8e-10 mbar - see attachment 1
open water inlet/outlet taps
NIM bin (MACBs, Ortec 480 Pulser, CAEN N1419ET) switched ON
USB-controlled ac mains relay switched ON - water & interlock LEDs active
Require VETAR2 (previous returned to Nik Kurz) and network switch (currently with AIDA@DESPEC)
Turbo pump ON - normal operation
Ion pumps ON - normal operation
16.30 ML informs us that ambient temperature of CRYRING cave has been increased from c. 16 deg C to c. 21 deg C and that this is likely the cause of the change in vacuum
pressure in YR09 and the rest of the storage ring.
obtained VETAR2 and crate
anydesk restarted |
|
|
|
|
|
|
|
|
|
|