Message ID: 402     Entry time: Wed May 24 08:35:12 2023
Author: JM, CB 
Subject: Strain reliefs (detector side) installed 

The new MACOR strain reliefs on the (detector side) have been installed. See attached photos

The two pieces  of MACOR are not fully closed but we are confident that the kapton wires are sufficiently clamped.

In order for the strain reliefs to be installed in the correct poisition, two nuts on the jacking plate of the cable harness had to be removed. This is as the groove in the strain relief was not deep enough to accomadate the nut. 

Attached is also a video of the wires (not) moving with the strain reliefs mounted