10.00 Outflow manifold returned from workshop with leaking attachment point welding re-worked.
11.30 Outflow hoses re-attached to outflow manifold. Faulty male Swagelok replaced.
Outflow manifold tested on floor - no leaks observed.
Mounted outflow manifold. All inflow/outflow hoses attached to FEE64s - see attachments 1-4.
FEE64 #11 detached adaptor PCB - temporarily secured by cable tie.
Inflow/outflow valves opened. Pin hole leak observed at 3/8" connector gasket - indicated by arrow attachment
3.
Leak fixed by tightening fitting.
1x spare inflow/outflow connected together - attachment 4.
11.40 CARME IE514 3.8e-11mbar, ambient temperature 16.5 deg C
15.05 8x AIDA FEE64 power cables ( #1 - #8 ) installed FEE64s #1 - #8
Note - Samtec lock clips broken on 3 of 8 cables due to careless handling/poor storage
AIDA FEE64 power cables tied to one of the FEE64 copper coolant pipes.
See attachments 5-7
15.30 USB-controlled ac mains relay - USB connected to RPi CARMEbakingpi currently used for CAEN N1419ET
Rear panel key switched to override - no flow/d.p. sensors currently installed - front panel water ( yellow ) and override ( red ) LEDs active.
16.15 Transfer CARMEbakingpi from CRYRING network to AIDA 24-port Gbit network switch connected to carme-gsi
CARMEbakingpi becomes nnrpi1 p- can connect OK via ssh from carme-gsi
Start server ( port 8015 ) manually
ssh pi@nnrpi1
/MIDAS/Tcl/Httpd/linux-arm/Tcl-Httpd-server &
See attachment 8 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|