Arrived at the control room to find we had lost connection to the power relay. We rebooted the nnrpi1 and re-connected according to the steps laid out in elog 471 -> all okay
11:00 Restarted daq with no changes to the system from yesterday. Bias consistent with previous days. All pulser peaks are slightly improved (even more for the pn pulser, one goes from 91 chn to 78). No pulser observed in nn FEE modules 9, 10, 13, 14 again.
ADC data items check OK - all ~3k
Rates - all OK - attachment 1
1.8L spectra. - attachment 2,3
1.8W waveforms - attachment 4,5
11:20 Changed shaping time from 0xf to 0x3 (2us)
ADC data items check OK - all ~3k attachment 6
Rates - all OK - attachment 7
1.8L pn spectra. - attachment 8
1.8W waveforms - attachment 9,10
11:35 Changed shaping time from 0x3 to 0x7 (4us)
ADC data items check OK - all ~3k attachment 11
Rates - all OK - attachment 12
1.8L pn spectra. - attachment 13
1.8W waveforms - attachment 14,15
11:45 Changed shaping time from 0x7 to 0xb (6us)
ADC data items check OK - all ~3k - attachment 16
Rates - all OK - attachment 17
1.8L spectra. - attachment 18,19
1.8W waveforms - attachment 20,21
11:59 Changed shaping time from 0xb to 0xd (7us)
ADC data items check OK - all ~3k - attchment 22
Rates - all OK - attachment 23
1.8L pn spectra. - attachment 24
1.8W waveforms - attachment 25, 26
12:09 Changed shaping time back to 0xf (8us)
ADC data items check OK - all ~3k - attchment 27
Rates - all OK - attachment 28
1.8L pn spectra. - attachment 29
1.8W waveforms - attachment 30,31
13:10 Power cycled FEE modules. nn pulser for FEEs 9,10,13,14 are now active again (no changes made in the cave) Pulser widths for nn and pn both consistent with before lunch
|