Message ID: 486
Entry time: Fri Feb 2 08:59:36 2024
|
Author: |
JM, TD, CB |
Subject: |
Run1 start - pulser widths spreadsheet |
|
|
Current pulser widths excel spreadsheet attached
09.58 DSSSD bias OK - attachment 1
Ambient temperature +16.1 deg C FEE64 temps OK - attachment 2
ADC data item stats OK - attachment 3
aida15 asic #2 high rate - asic u/s
system wide tests OK *except* aida09 - attachments 4-5
aida09 WR status 0x20 aida09 WR timestamp incorrect cf. all other FEE64s
per FEE64 Rate spectra - attachment 6
per FEE64 1.8.L spectra - attachments 7-8
aida14 pulse peak witdh 58 ch FWHM
aida11 pulse peak witdh 48 ch FWHM
per FEE64 1.8.W spetra - 20us FSR - attachments 9-10
12.00
aida09 WR timestamp issue resolved by FEE64 reboot
14.10
Safety pins removed from detector motion system. CB performs one, complete DSSSD movement cycle. Zero all histogram and stats per FEE64 Stat spectra ( log scale ) and
ADC
data item stats for the complete DSSSD movement cycle - attachments 11-12
No change observed in DSSSD leakage currents during DSSSD movement cycle.
15.35
Resolved issue from previous elog. Begin background run writing data to disk.
ASIC check load
Start DAQ G22-88/R1 background alpha
EG&G Ortec 448 Pulser OFF
ASIC settings 2024Jan20-12.02.08
slow comparator 0x64 ( all FEE64s *except aida15 asic #2 0xff )
All histograms set to zero
DSSSD bias OK - attachment 13
FEE64 temps OK - attachment 14
WR timestamps OK - attachment 15
ADC data item stats - attachment 16
aida15 asic #2 u/s - rate c. 250k
ADC data items/s Merger, Tape Server & Merger stats - attachments 17-19
data rate to disk c. 2Mb/s dominated by aida15 asic 2 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|