06:45
All system checks OK
Target density 3.4e11
11:50
System checks OK.
No beam in the ring yet. Althoug some injection may be taking place at low intensities
Target density appears much more stable than before.
Density currently just above 1e11 at/cm2
Note maintanace is required to appc025 so anydesk is currently down.
13:13
Merger OK, Tape server OK
ADC data item statistics looks as before. OK
FEE temperatures look OK
System checks OK
Leakage current stable
Attachments 1-5 (also waveforms)
15:00
Merger OK, Tape server OK
ADC data item statistics DO NOT looks as before. NOT OK
FEE temperatures look OK
System checks OK
Leakage current stable
Attachments 6-8
15:16
Notice the statistics for aida01-aida04 (DDSD 0, top left) were counting extremely high in the previous check (attachment 6). Previous rate was 0-1 counts.
We tried ASIC check load three times but not effect on the rate.
We then tried 'shaking the detectors' by moving the left pneumatic in and out again.
Rates go back to normal levels (0-1 counts)
|